Vitros recall
VITROS (R) Chemistry Products Calibrator Kit 17, Product Code/ Catalog Number 6801701 -- For in vitro diagnostic use only --- VITROS Chemistry…
Why was Vitros recalled?
Calibrator values assigned for the calibrator kit lots were released using a reference method that was undergoing validation for a new lot of European Reference Material (ERM). The validation process was not complete prior to calibrator value assignment for VITROS (R) Chemistry Products hsCRP Reagent (Product Code 6801739), Generation (GEN) 28.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lots 1732 (exp. 31-Jul-2014), 1743 (exp. 31-Jul-2014), 1753 (exp. 28-Feb-2015), 1764 (31-Jul-2015)
- Product
- VITROS (R) Chemistry Products Calibrator Kit 17, Product Code/ Catalog Number 6801701 -- For in vitro diagnostic use only --- VITROS Chemistry…
- Recalling firm
- Ortho-Clinical Diagnostics
- Quantity
- Domestic - 612 units, Foreign - 461 units
- Distribution
- Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
- Recall date
- June 25, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)