FDAClass IApril 22, 2026

B. Braun Hemodialysis Bloodlines recall

B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG. Model Number: SL-2010M2096

Why was B. Braun Hemodialysis Bloodlines recalled?

Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model Number: SL-2010M2096; UDI-DI Primary: 04046964367786; UDI-DI Unit: 04046964367779; All lots manufactured since 25MAY2025.

Product
B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG. Model Number: SL-2010M2096
Recalling firm
B Braun Medical Inc
Quantity
2,653,711 units
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Recall date
April 22, 2026
Classification
Class I
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls