FDAClass IIJune 16, 2021

Ingenia Ambition S recall

Ingenia Ambition S - Model Number: 781359 Diagnostic MR system.

Why was Ingenia Ambition S recalled?

The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities. A mechanical issue was detected on these hoisting interface blocks on top of the magnet.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

REF 781359/ UDI-DI 00884838090057 Serial numbers: 48000 48001 48003 48004 48009 48005 48015 48017 48020 48011 48022 48012 48014 48028 48026 48037 48023 48043 48019 48160 48144 48180 48112 48052 ***Updated November 2023*** Added Serial Numbers 48051 48010 48038 48018 48039 48027 48122 48046 48036 48047 48133 48056 48169 48148 48154 48055 48128 48194 48173 48134 48040 48155 48186 48035 48181 48188 48198

Product
Ingenia Ambition S - Model Number: 781359 Diagnostic MR system.
Recalling firm
Philips North America Llc
Quantity
156 MR systems
Distribution
Domestic: FL, GA, MI, MO, NH, NJ, NY, PA, SC, TX, VT. Foreign: Australia, Austria, Bangladesh, Belgium, Brazil, Denmark, Finland, France, Germany, Iraq, Italy, Japan, Libya, Moldova, Netherlands, Oman, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom.
Recall date
June 16, 2021
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls