Axiom Aristos MX Solid State Flat Panel Digital Imager recall
Axiom Aristos MX Solid State Flat Panel Digital Imager, Model number 5895003. The device is used for diagnostic X-ray imaging.
Why was Axiom Aristos MX Solid State Flat Panel Digital Imager recalled?
Siemens issued a Field Safety Notice about the potential hazard to patients or operators when using the function "table top float" of the AXIOM Aristos MX. When repositioning a patient lying on the table top and releasing the table top, it is possible that the fingers of the patient or operator are between the table top and the detector tray and if the table top is moved, the fingers can be squee
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model Number 5895003 with serial numbers 1052 through 2500.
- Product
- Axiom Aristos MX Solid State Flat Panel Digital Imager, Model number 5895003. The device is used for diagnostic X-ray imaging.
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 401
- Distribution
- Nationwide Distribution.
- Recall date
- July 31, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)