FDAClass IIApril 29, 2020

Siemens N Latex CDT Kit- IVD quantitative determination recall

Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. Siemens Material Number (SMN)…

Why was Siemens N Latex CDT Kit- IVD quantitative determination recalled?

Negative bias for the carbohydrate-deficient transferrin (CDT) measurement in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

US Lot number: 49822 UDI: 842768018534 OUS: Lot numbers: 49729, 49849, 49985, 50084 Exp. Date: 2020-04-11

Product
Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. Siemens Material Number (SMN)…
Recalling firm
CSL Behring GmbH
Quantity
297 PCK US; 1247 PCK OUS
Distribution
IN, KS, MO, OH, VA Foreign: Australia, Austria, Bangladesh, Belgium, Canada, China Czech Republic, Egypt, Finland, France, Germany, Great Britain, Hungary India, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Republic Korea, Russia Federation, Slovakia, Slovenia, South Africa, Spain Sweden, Switzerland, Taiwan
Recall date
April 29, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls