FDAClass IIApril 29, 2020

Temperature Sensing 100% Silicone Foley Catheter recall

Temperature Sensing 100% Silicone Foley Catheter, Part Number 102202101863DO - Product Usage: is a 3 lumen silicone catheter with a sensor to measure…

Why was Temperature Sensing 100% Silicone Foley Catheter recalled?

Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Numbers: V19012401 V19012403

Product
Temperature Sensing 100% Silicone Foley Catheter, Part Number 102202101863DO - Product Usage: is a 3 lumen silicone catheter with a sensor to measure…
Recalling firm
Degania Medical Devices Pvt. Ltd.
Quantity
2698
Distribution
Worldwide distribution - US Nationwide distribution including state of IL and the country of Mexico.
Recall date
April 29, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)