FDAClass IIMay 15, 2024

Intelli-C: Right side suspended Tabletop recall

Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system

Why was Intelli-C: Right side suspended Tabletop recalled?

X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

REF/UDI-DI/Serial Numbers: 03400000/05713464000527/627002, 627004, 627007, 639001, 639002, 639003, 639004, 639007, 639010, 647001, 647004, 647005, 648001, 651007, 658002, 658003, 658006, 658010, 666008, 675001, 675004, 682006, 682010, 682011, 696003, 696004, 696006, 696007, 696012. 03400010/05713464000121/647009, 651002, 651006, 666009, 682003, 682023

Product
Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system
Recalling firm
NRT X-RAY A/S
Quantity
35
Distribution
US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, WV, LA, NY, PA, HI, NV, WI, MN, MI.
Recall date
May 15, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls