Refinity Short Tip Rotational Intravascular Ultrasound recall
Refinity Short Tip Rotational Intravascular Ultrasound (IVUS) Catheter; Product number: 400-0200.273; Catalog number 89900. Cardiovascular: The…
Why was Refinity Short Tip Rotational Intravascular Ultrasound recalled?
High complaint rates of kinking near the junction of the proximal shaft to telescope section of the catheter. Kinking could result in disruption of normal clinical workflow or loss of imaging.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot # 50069132, Expiry date 6/30/2018; Lot # 50083527, Expiry date 11/30/2018; Lot # 50083533, Expiry Date 11/30/2018.
- Product
- Refinity Short Tip Rotational Intravascular Ultrasound (IVUS) Catheter; Product number: 400-0200.273; Catalog number 89900. Cardiovascular: The…
- Recalling firm
- Volcano Corporation
- Quantity
- 718 total devices
- Distribution
- Distribution to US and Japan
- Recall date
- April 19, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)