FDAClass IIJune 9, 2021

Synapse Cardiovascular recall: Software Issue

Synapse Cardiovascular (Synapse CV). Software Versions: 6.0 to 6.2.1 The Synapse Cardiovascular image is a management and reporting system

Why was Synapse Cardiovascular recalled?

Software versions 6.0.4 to 6.2.1 using Advanced Reporting -possibility that a previously assigned internal patient database ID can be reused for a new patient. It was discovered that this value does not increase when a patient merge activity is executed in a specific sequence

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software Versions: 6.0 to 6.2.1

Product
Synapse Cardiovascular (Synapse CV). Software Versions: 6.0 to 6.2.1 The Synapse Cardiovascular image is a management and reporting system
Recalling firm
Fujifilm Medical Systems U.S.A., Inc.
Quantity
44 US and 4 OUS
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Bermuda, Israel, Hadassah Medical Center.
Recall date
June 9, 2021
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls