Kappa recall
Kappa, Catalog Number: MS18853 in combination with Scio, Scio Four, Scio Four Oxi plus, Scio Four Oxi, Scio Four plus. Catalog Number(s)…
Why was Kappa recalled?
It was reported that a set low O2 alarm does not go off although the measured O2 level is below the alarm limit.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
VF10.0 software
- Product
- Kappa, Catalog Number: MS18853 in combination with Scio, Scio Four, Scio Four Oxi plus, Scio Four Oxi, Scio Four plus. Catalog Number(s)…
- Recalling firm
- Draegar Medical Systems, Inc.
- Quantity
- 2,156 (total for all devices)
- Distribution
- Worldwide - US Nationwide distribution in the states of: AL, CA, GA, IL, LA, MA NY, OH, PA, TN, TX and VA and the countries of: AR AT BE BG BN BO BY CA CH CL CO CZ DE EG ES FI FR GB GR HR HU ID IE IL IN IT LT MX MY NL NO NP PA PE PK PL PT RO RS RU SA SE SK TH TR TW UG UZ VN ZA
- Recall date
- April 19, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)