Boston Scientific Cardiac Resynchronization Therapy recall: Software Issue
Boston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL, Model Number U225; 4. VISIONIST CRT-P EL, Model Number U226; 5. VISIONIST CRT-P EL MRI, Model Number U228.
Why was Boston Scientific Cardiac Resynchronization Therapy recalled?
Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
The advisory population includes all models listed; however, the bounding differs by battery type: " All serialized DR-EL pacemakers and CRT-Ps from the ACCOLADE family are included in the advisory population. " ACCOLADE DR-SL and SR-SLs with a use-by-date (UBD) on or before 30 June 2025 are included in the advisory population. Model number alone will not precisely identify individual DR-SL or SR-SL devices in the advisory population. To determine if a device is affected, enter a model/serial into the device lookup tool at www.BostonScientific.com/lookup.
- Product
- Boston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL, Model Number U225; 4. VISIONIST CRT-P EL, Model Number U226; 5. VISIONIST CRT-P EL MRI, Model Number U228.
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 685,776 units
- Distribution
- worldwide
- Recall date
- May 6, 2026
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)