FDAClass IISeptember 28, 2022

Aquac UNO H [120v recall

AquaC UNO H [120V, 60Hz, 1700VA, US-Version- a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-0970-1…

Why was Aquac UNO H [120v recalled?

Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI: (01)4251285200234(11)200422(21)0CHS3987 (01)4251285200234(11)200526(21)0CHS4108 (01)4251285200234(11)210623(21)1CHS5372 (01)4251285200234(11)210707(21)1CHS5405 (01)4251285200234(11)190305(21)9CHS2881 (01)4251285200234(11)190529(21)9CHS3090 Serial Numbers: 0CHS3987 0CHS4108 1CHS5372 1CHS5405 9CHS2881 9CHS3090

Product
AquaC UNO H [120V, 60Hz, 1700VA, US-Version- a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-0970-1…
Recalling firm
Fresenius Medical Care Holdings, Inc.
Quantity
6 units
Distribution
US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV, NY, PA, TX.
Recall date
September 28, 2022
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls