Aquac UNO H [120v recall
AquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-10VA-1…
Why was Aquac UNO H [120v recalled?
Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI: (01)4251285200234(11)210628(21)1CHS5379 (01)4251285200234(11)160927(21)6CHS1270 (01)4251285200234(11)161109(21)6CHS1394 (01)4251285200234(11)161109(21)6CHS1435 (01)4251285200234(11)170131(21)7CHS1569 (01)4251285200234(11)170201(21)7CHS1583 (01)4251285200234(11)170213(21)7CHS1619 (01)4251285200234(11)170314(21)7CHS1641 (01)4251285200234(11)170519(21)7CHS1790 (01)4251285200234(11)170807(21)7CHS1953 (01)4251285200234(11)180326(21)8CHS2197 (01)4251285200234(11)181010(21)8CHS2608 (01)4251285200234(11)181102(21)8CHS2633 (01)4251285200234(11)181107(21)8CHS2644 Serial Numbers: 1CHS5379 6CHS1192 6CHS1194 6CHS1195 6CHS1196 6CHS1197 6CHS1198 6CHS1199 6CHS1200 6CHS1201 6CHS1233 6CHS1270 6CHS1394 6CHS1435 7CHS1569 7CHS1583 7CHS1619 7CHS1641 7CHS1790 7CHS1953 8CHS2197 8CHS2608 8CHS2633 8CHS2644
- Product
- AquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-10VA-1…
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Quantity
- 24 units
- Distribution
- US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV, NY, PA, TX.
- Recall date
- September 28, 2022
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)