FDAClass IIMay 15, 2024

Duraflow 2 recall: Foreign Material

DURAFLOW 2, 32CM STRAIGHT BASIC KIT PG-Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number…

Why was Duraflow 2 recalled?

Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: 25051684014752 Lot Numbers: 5805839

Product
DURAFLOW 2, 32CM STRAIGHT BASIC KIT PG-Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number…
Recalling firm
Angiodynamics, Inc.
Quantity
35 units
Distribution
VA
Recall date
May 15, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)