Advance Enforcer 35 Focal Force PTA Balloon Catheter recall
Advance Enforcer 35 Focal Force PTA Balloon Catheter 6mm x 4cm, 5FR/50cm, Catalog Number ASB5-35-50-6-4, REF Number G35248 Product Usage: The Advance…
Why was Advance Enforcer 35 Focal Force PTA Balloon Catheter recalled?
Multiple complaints were received for balloons bursting below the rated burst pressure. Potential adverse events that may occur if an affected product is used include a delay in the procedure, additional intervention, vessel injury, and balloon fragmentation in the patient.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Numbers: 9234424, 9331618
- Product
- Advance Enforcer 35 Focal Force PTA Balloon Catheter 6mm x 4cm, 5FR/50cm, Catalog Number ASB5-35-50-6-4, REF Number G35248 Product Usage: The Advance…
- Recalling firm
- Cook Inc.
- Quantity
- 34
- Distribution
- Worldwide Distribution - US Nationwide The affected product was distributed to the following states: FL, OH, WA. The products were distributed to the following foreign countries: Belgium, Denmark, Finland, Germany, Italy, Netherlands, Norway, Poland, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
- Recall date
- June 26, 2019
- Classification
- Class I
- Status
- Terminated
- Agency
- FDA (device)