FDAClass IIJuly 24, 2013

Hydra Vision DR 60/80 System recall

Hydra Vision DR 60/80 System, Models 700539 and 700540, Liebel-Flarshiem Company. The System is used in radiologic imaging procedures.

Why was Hydra Vision DR 60/80 System recalled?

A complaint was received in which a hospital physicist indicated the dosage was too high on the HUT system. During the service call the systems Maximum Fluoro Dose was found to be 30 R/min which is three times the Maximum Fluoro Dose limit of 10 R/min.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

The following catalog numbers, serial numbers and associated manufacturing dates are affected by this recall: 700539 CI0807H273 August2007; 700539 CI0106H162 January 2006; 700539 CI1010H523 October 2010; 700539 CI0811H567 August 2011; 700539 CI0108H313 January 2008; 700539 CI0309H418 March 2009; 700539 15031229 Apri12003; 700539 14031227 Apri12003; 700539 CI0404H045 January 2005; 700539 CI0504H054 May 2004; 700539 CI1106H208 November 2006; 700539 CI0509H424 May 2009

Product
Hydra Vision DR 60/80 System, Models 700539 and 700540, Liebel-Flarshiem Company. The System is used in radiologic imaging procedures.
Recalling firm
Leibel-Flarsheim Company LLC
Quantity
12 units
Distribution
Nationwide Distribution (US), including the states of TN, KY, CO, KS, CT, HI, and AZ.
Recall date
July 24, 2013
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls