ExactaMix 1200 and 2400 Compounders. Used as an recall
ExactaMix 1200 and 2400 Compounders. Used as an automated pumping system that compounds multiple sterile ingredients into a finished solution.
Why was ExactaMix 1200 and 2400 Compounders. Used as an recalled?
Baxter Healthcare is recalling certain ExactaMix automated compounding devices due to the complementary metal oxide semi-conductor battery failure resulting in resetting the internal clock of the device to midnight December 31, 2010.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
ExactaMix 1200 serial numbers 45004-45481 (sequentially), and ExactaMix 2400 serial numbers 40005-40335.
- Product
- ExactaMix 1200 and 2400 Compounders. Used as an automated pumping system that compounds multiple sterile ingredients into a finished solution.
- Recalling firm
- Baxa Corporation
- Quantity
- 753 units
- Distribution
- Worldwide Distribution-USA (nationwide) including Puerto Rico, and the countries of Canada, Latin America, Europe/Middle East/Africa, and Taiwan.
- Recall date
- July 24, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)