Synchron Systems Opiate 2000 ng Reagent recall
Synchron Systems Opiate 2000 ng Reagent (OP2 1 x 250); catalogue no. 475006 for use in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON…
Why was Synchron Systems Opiate 2000 ng Reagent recalled?
Cross-reactivity testing may not meet the claims indicated in the Opiate 2000 ng (OP2) Instructions For Use. The presence of certain compounds in urine may result in false positive or false negative patient test results.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog Number 475006 ; UDI: 15099590222031; Lot Numbers: M904147 and older including M810039, M901087 and M904147) and 4 lots OP2 (M805122, M810039, and M901087. .
- Product
- Synchron Systems Opiate 2000 ng Reagent (OP2 1 x 250); catalogue no. 475006 for use in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON…
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 1610 units
- Distribution
- Distributed throughout the United States and to Puerto Rico. States include: CA, NC, OR, MI, GE, AL, SC, TX, NJ, IA, MS, IN, MA, NC, OH, AZ, LA, MA, NY, PA, KY, WV, MI, WA, CO, IL, VA, FL, KS, MD, ID, DE, NV, OK, AR, HI, WI, NY.
- Recall date
- April 29, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)