FDAClass IISeptember 21, 2022

INOmax DSIR -(delivery system) delivers INOMAX recall: Software Issue

INOmax DSIR -(delivery system) delivers INOMAX (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in…

Why was INOmax DSIR -(delivery system) delivers INOMAX recalled?

The INOmax DSIR Delivery System, manufactured by Mallinckrodt Pharmaceuticals, is not compatible with the upgraded version of the Leoni Plus ventilator, manufactured by Lowenstein Medical. The Leoni Plus ventilator was recently upgraded in 2021 by Lowenstein Medical, with new hardware and software, and the upgraded version cannot be appropriately connected with the INOmax DSIR as instructed in the INOmax DSIR Operators Manual (OM). The INOmax DSIR OM requires the use of a one-way valve with the Leoni Plus ventilator during High Frequency Oscillatory Ventilation mode and warns that use of the device without a one-way valve may result in high Nitric Oxide (NO) delivery.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All serial numbers

Product
INOmax DSIR -(delivery system) delivers INOMAX (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in…
Recalling firm
Mallinckrodt, LLC.
Quantity
4,225 units OUS
Distribution
Foreign Only: Argentina Australia Austria Beligum Canada Chile Columbia Czech Republic Denmark Ecuador Estonia Finland France Germany Great Britain Hungary Iceland Ireland Italy Latvia Mexico Netherlands Norway Poland Portugal Romania Slovakia South Africa Spain Sweden Switzerland UAE Uruguay
Recall date
September 21, 2022
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls