FDAClass IISeptember 21, 2022

Pinnacle3 Radiation Therapy Planning System recall

Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.

Why was Pinnacle3 Radiation Therapy Planning System recalled?

When computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model 870231: Software version 16.0.4, UDI (01)00884838099081(10)16.0.4.60330; Software version 16.4.2, UDI (01)0084838103054(10)16.4.2.60330, UDI (01)00884838102965(10)16.4.2.60330, UDI (01)00884838102972(10)16.4.2.60315, UDI (01)00884838102989(10)16.4.2.60330; UDI (01)00884838103191(10)16.4.2.60330; and Software version 16.4.3, UDI (01)00884838102965(10)16.4.3.60360. Model 870237: Software version 16.4.6, UDI (01)00884838103566(10)16.4.6.21197; and Software version 18.0.1, UDI (01)00884838103566(10)18.0.1.2149 and (01)00884838103566(10)18.0.1.21197.

Product
Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.
Recalling firm
Philips Medical Systems (Cleveland) Inc
Quantity
169 systems
Distribution
Distribution was made to CA, FL, GA, IA, IL, MA, MD, ME, MI, MN, NJ, NM, NV, NY, OH, OK, RI, SC, TN, TX, and WA. There was government distribution and no military distribution. Foreign distribution was made to Australia, Austria, Belgium, China, France, Germany, Italy, Japan, Jordan, Lebanon, Poland, Portugal, Romania, Russian Federation, Singapore, Spain, Switzerland, Taiwan, Thailand, and United Kingdom.
Recall date
September 21, 2022
Classification
Class II
Status
Ongoing
Agency
FDA (device)