FDAClass IJune 28, 2023

Crt-d Dtbb1q1 Viva Quad S Is4/df1 US recall

CRT-D DTBB1Q1 VIVA QUAD S IS4/DF1 US, Model Number DTBB1Q1; Implantable Cardioverter Defibrillators

Why was Crt-d Dtbb1q1 Viva Quad S Is4/df1 US recalled?

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

GTIN 00643169720206, Lot Serial Numbers: BLM202912H, BLM203000H, BLM202895H, BLM202965H, BLM203007H, BLM203028H, BLM203031H, BLM203063H, BLM202844H, BLM202900H, BLM203042H, BLM203014H, BLM203015H, BLM203062H, BLM202811H, BLM202838H, BLM202857H, BLM202858H, BLM202881H, BLM202923H, BLM203001H, BLM203021H, BLM202848H, BLM202878H, BLM202827H, BLM202952H, BLM203061H, BLM202909H, BLM202921H, BLM202942H, BLM202830H, BLM202850H, BLM202886H, BLM202888H, BLM202829H, BLM202911H, BLM202958H, BLM202982H, BLM203037H, BLM202945H, BLM202929H, BLM202949H, BLM202973H, BLM202828H, BLM202990H, BLM202947H, BLM202960H, BLM202854H, BLM202944H, BLM202810H, BLM202847H, BLM202862H, BLM202880H, BLM202904H, BLM202946H, BLM202974H, BLM202979H, BLM203006H, BLM203012H, BLM203040H, BLM202932H, BLM202913H, BLM202940H, BLM202842H, BLM202980H, BLM203029H, BLM203058H, BLM203065H, BLM202972H, BLM203051H, BLM202867H, BLM202948H, BLM203004H, BLM202832H, BLM202999H, BLM202885H, BLM202976H, BLM203035H, BLM203009H, BLM203024H, BLM202914H, BLM203002H, BLM202833H, BLM202983H, BLM202985H, BLM202986H, BLM202987H, BLM202988H, BLM202992H, BLM202996H, BLM203003H, BLM203005H, BLM203008H, BLM203016H, BLM203022H, BLM203043H, BLM203046H, BLM203047H, BLM203048H, BLM203049H, BLM203066H, BLM202851H, BLM202852H, BLM202866H, BLM202875H, BLM202984H, BLM203026H, BLM203054H, BLM202860H, BLM202927H, BLM202846H, BLM202943H, BLM202951H, BLM202989H, BLM202876H, BLM202825H, BLM202915H, BLM202954H, BLM202849H, BLM202836H, BLM203053H, BLM203064H, BLM203067H

Product
CRT-D DTBB1Q1 VIVA QUAD S IS4/DF1 US, Model Number DTBB1Q1; Implantable Cardioverter Defibrillators
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Distribution
US Nationwide - Worldwide Distribution
Recall date
June 28, 2023
Classification
Class I
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls