FDAClass IIIJuly 24, 2013

TDHistology/Cytology Classification name recall: Software Issue

TDHistology/Cytology Classification name: Calculator/data processing module for clinical use (862.2100). Intended use: Laboratory Information System…

Why was TDHistology/Cytology Classification name recalled?

A recall was initiated by Technidata because use of symbols (less than or equal to or greater than or equal to or any other symbol) in Microsoft Word reports when using electronic transmissions of results by of HPRIM (including HPRIM Medecin via Kermit) or HL7 must be absolutely prohibited. Such symbols are not supported by the Microsoft Word interface used by the Histology/Cytology software and

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All TDHistology/Cytology product versions.

Product
TDHistology/Cytology Classification name: Calculator/data processing module for clinical use (862.2100). Intended use: Laboratory Information System…
Recalling firm
Technidata S.A.
Quantity
1 site
Distribution
US Distribution CA only.
Recall date
July 24, 2013
Classification
Class III
Status
Terminated
Agency
FDA (device)