FDAClass IIApril 8, 2026

MiniMed recall

MiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)

Why was MiniMed recalled?

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 780G Insulin Pump: 763000411541, 763000441449, 763000523305, 763000521530, 763000521516, 763000533861, 763000544317, 763000734459, 763000758530, 763000758547, 763000578688, 763000734435, 763000504762, 763000734466, 763000924102, 763000924089, 763000454296, 763000854829, 763000411411, 763000411503, 763000384272, 763000411466, 763000411428, 763000411572, 763000411589, 763000411558, 763000384265, 763000411473, 763000411374, 763000411435, 763000411404, 763000384289, 763000411459, 763000411510, 763000411626, 763000411565, 763000411527, 763000411619, 763000411596, 763000411497, 763000411398, 763000411534, 763000411381, 763000384258, 763000411480, 763000411602, 763000411640, 763000411442, 763000441432, 763000442194, 763000521523, 763000442187, 763000521509, 763000521547, 763000578671, 763000596194, 763000639228, 763000639235, 763000639266, 763000974381, 763000974374, 763000921750, 763000921767, 763000921590, 763000921637, 763000933593, 763000891886, 763000921729, 763000921866, 763000921781, 763000921811, 763000921576, 763000921583, 763000921620, 763000921859, 763000891909, 763000921743, 763000944162, 763000921712, 763000921668, 763000933579, 763000933555, 763000933586, 763000933609, 763000921613, 763000933715, 763000933548, 763000944209, 763000891893, 763000933623, 763000933722, 763000933814, 763000933708, 763000933777, 763000933685, 763000933531, 763000933678, 763000933821, 763000933746, 763000933562, 763000944193, 763000533878, 763000734442, 763000854942, 763000914127: Medtronic MiniMed 780G with Instinct sensor System User Guide(M018856C024 2, M018857C035 2); Medtronic MiniMed 780G with Simplera Sync sensor System User Guide(M018857C034 2); Medtronic MiniMed 780G with Guardian 4 Sensor System User Guide(M028790C004 2, M028792C004 2); Medtronic MiniMed 780G with Guardian 4 Sensor System User Guide(M028792C015 1, M028792C001 3, M028792C002 1, M028792C006 1, M028792C003 1, M028792C005 1); Medtronic MiniMed 780G with Instinct sensor System User Guide(M018857C008 1, M018857C011 1, M018857C029 1, M018856C022 1, M018857C013 1, M018856C007 1, M018857C001 4, M018856C001 4, M018856C010 1, M018857C012 1, M018856C020 1, M018857C009 1, M018857C032 1, M018857C022 1, M018856C021 1, M018856C013 1, M018857C028 1, M018857C033 1, M018856C017 1, M018857C016 1, M018856C018 1); Medtronic MiniMed 780G with Simplera Sync sensor System User Guide(M018856C023 2, M018857C034 2); MiniMed 780G System Technical Guide(M075489C003 1, M075489C002 1, M075489C001 2); MiniMed 780G System User Guide(M072481C001 2); MiniMed 780G with Guardian 4 Sensor System User Guide mg(, M073070C003 1, M073070C001 2, M073070C002 1); MiniMed 780G with Guardian 4 Sensor System User Guide mm(M076019C002 1, M076019C001 1); MiniMed 780G with Guardian 4 Sensor System User Guide(M028790C003 1, M028790C002 1); MiniMed 780G system Changing the Medtronic Extended Infusion Set with Extended Reservoir(M048911C001 2, M048911C003 1, M048911C002 1); MiniMed 780G system Changing the MiniMed Mio Advance Infusion Set and Reservoir(M047607C001 1); MiniMed 780G system Changing the MiniMed Silhouette Infusion Set and Reservoir(M047610C001 1, M047610C003 1, M047610C002 1); MiniMed 780G system Changing the MiniMed Sure-T Infusion Set and Reservoir(M047609C001 1, M047609C003 1, M047607C003 1, M047607C002 1, M047609C002 1); MiniMed 780G System User Guide(M024832C003 2, M024829C004 2, M024832C004 2, M024829C005 2); MiniMed 780G System User Guide mg/dl(M014786C005 2, M014784C005 2, M014786C007 1, M014784C007 1, M014784C004 1, M014786C004 1, M014786C008 1, M014784C008 1, M014784C002 1, M014786C006 1, M014784C006 1, M014784C016 1, M014786C016 1, M014784C012 1, M014786C012 1, M014786C002 1, M014784C010 1, M014786C010 1, M014784C003 1, M014786C003 1, M014784C011 1, M014786C011 1, M014786C013 1, M014784C013 1, M014784C015 1, M014786C015 1, M014786C019 1, M014784C019 1); MiniMed 780G System User Guide mmol/L(M014787C002 1, M014785C002 1, M014785C007 2, M014787C007 1, M014787C005 1, M014785C005 1, M014787C006 1, M014785C006 1, M014785C023 2, M014787C023 2, M014787C010 1, M014785C010 1, M014787C011 1, M014785C011 1, M014787C021 1, M014785C021 1, M014787C008 1, M014785C008 1, M014785C009 1, M014787C009 1, M014787C022 1, M014785C022 1, M014785C027 1, M014787C027 1, M014787C004 2, M014785C004 2, M014785C003 1, M014787C003 1, M014785C030 1, M014787C030 1, M014785C013 2, M014787C013 2, M014785C019 1, M014787C019 1, M014785C020 1, M014787C020 1, M014787C018 1, M014785C018 1, M014787C012 1, M014785C012 1, M014787C016 1, M014785C016 1, M014787C024 1, M014785C024 1, M014785C014 2, M014787C014 1, M014787C028 1, M014785C028 1); MiniMed 780G with Guardian 4 Sensor System User Guide(M060264C012 1, M060263C011 1, M060264C018 1, M060263C015 1, M060264C017 1, M060263C014 1, M060264C006 1, M060264C004 1, M060263C001 3, M060263C016 1, M060264C007 1, M060264C008 1, M060264C009 1, M060264C010 1, M060264C011 1, M060264C001 3, M060264C013 1, M060264C014 1, M060264C023 1, M028790C005 1, M060263C020 1, M060264C015 1, M060264C016 1, M060264C020 1, M060263C021 1, M028790C006 1, M060263C022 1, M056377C001 3, M056377C003 1, M056377C002 1, M060263C023 1, M060264C021 1, M060264C022 1, M060263C024 1); MiniMed 780G with Guardian 4 Sensor System User Guide mg/dl(M028790C001 3, M028790C011 1, M028790C007 1, M028790C008 1, M028790C010 1, M028790C019 1); MiniMed 780G with Guardian 4 Sensor System User Guide mmol/L(M028792C012 1, M028792C013 1, M028792C007 1, M028792C011 1, M028792C025 1, M028792C014 1, M028792C001 3, M028792C033 1, M028792C010 1); MiniMed 780G with Instinct sensor System User Guide(M018857C004 1); MiniMed 780G with Simplera Sync sensor System User Guide(M060262C008 1, M060261C002 1, M018857C003 1, M018856C003 1, M060262C015 1, M018856C005 1, M060261C004 1, M060261C003 1, M018856C006 1, M060262C014 1, M060262C002 1, M060261C001 3, M060261C005 1, M060262C003 1, M060262C004 1, M060262C005 1, M060262C006 1, M060262C007 1, M060262C001 3, M060262C009 1, M018857C015 1, M060262C010 1, M018856C009 1, M060261C009 1, M060262C021 1, M018857C034 2, M060262C012 1, M060262C013 1, M060262C016 1, M018856C012 1, M060261C010 1, M060261C011 1, M056047C001 3, M056047C003 1, M056047C002 1, M060261C012 1, M060262C019 1, M060262C020 1, M060261C013 1); MiniMed 780G with the Guardian 4 Sensor System User Guide(M036461C004 2, M036461C001 2, M036461C007 1, M036461C003 1, M036461C002 1, M036461C006 1); MiniMed 780G with the Guardian Sensor (3) System User Guide(M010682C005 2, M010682C008 1); MiniMed 780G with the Guardian Sensor 3 System User Guide(M010682C002 2, M010682C004 1, M010682C007 1, M010682C003 1)

Product
MiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
Recalling firm
Medtronic MiniMed, Inc.
Quantity
453,144
Distribution
Worldwide - US Nationwide distribution including in the states of OR, CA, AZ, NY,TX,FL, OH, MO, TN,NJ, NC, SC, IL, VA, WA, DC, GA, PA, MS,IA, MI, NH, OK, SD, MT, WI, WV, MA, LA, MN, RI, AR, ME, AL, IN, UT, KY, CO, ND, MD, CT, NM, KS, ID, NV, DE, NE, VT, AK, HI, WY, PR, VI and the countries of Taiwan, New Zealand, Australia, Bangladesh, Brunei Darussalam, Cambodia, Indonesia, Malaysia, Maldives, Philippines, Thailand, Vietnam, Singapore, Pakistan, Hong Kong, China, Argentina, Aruba, Barbados, Bermuda, Bolivia, Plurinational State Of, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, Puerto Rico, Trinidad And Tobago, Uruguay, Venezuela, Canada, South Korea, India, Albania, Algeria, Andorra, Armenia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Bulgaria, Canary Islands, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guyana, French Polynes., Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Macedonia, Malawi, Malta, Martinique, Mauritius, Mayotte, Moldavia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Qatar, Reunion, Romania, Russia, San Marino, Saudi Arabia, Serbia, Sierra Leone, Slovakia, Slovenia, South Africa, Spain, St.Pier,Miquel., Sudan, Sweden, Switzerland, Syrian Arab Republic, Tanzania, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Yemen, Zambia, Zimbabwe.
Recall date
April 8, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls