FDAClass IIJuly 24, 2013

CORE Universal Driver recall: Software Issue

CORE Universal Driver REF 5400- 99 Rx Only This drill may also be used with theTotal Performance System (TPS").The Stryker Total Performance System…

Why was CORE Universal Driver recalled?

The safety margin values detailed in the Engineering Design were entered incorrectly into the programming software. Handpieces that have been programmed with the incorrect parameters could result in unintended activation. There is a potential for injury to the patient or user as a result of unintended forward or reverse activation. There is also the potential that the handpiece may not reach 100

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Number 0326801633, 0330802913, 0406200763, 0410401383, 0410600013, 0534203093,0627600783, 0730101013, 0730101033, 0805000553, 0809829593, 0935603083,1011903103, 1011903493, 1013814003, 1025400463, 1025400643, 1126901343, and 1128022593

Product
CORE Universal Driver REF 5400- 99 Rx Only This drill may also be used with theTotal Performance System (TPS").The Stryker Total Performance System…
Recalling firm
Stryker Instruments Div. of Stryker Corporation
Quantity
11
Distribution
Worldwide distribution: US (nationwide) including states of: AZ, FL, GA, NC, NY, SC, TX and VA; and countries of: India, Netherlands and Japan.
Recall date
July 24, 2013
Classification
Class II
Status
Terminated
Agency
FDA (device)