FDAClass IIJuly 24, 2013

Eon Mini - 3721 recall

Eon Mini - 3721 (Charger). Product Usage: Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or…

Why was Eon Mini - 3721 recalled?

St. Jude Medical has informed your doctor that a number of cases have been reported in which discomfort associated with heating occurred at the device site while patients were using the Charging System to charge their spinal cord stimulator. In three cases, patients received a burn to the skin (one second-degree and two first-degree burns) at the implant site. As of June 30, 2012, St. Jude Medica

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All serial numbers manufactured

Product
Eon Mini - 3721 (Charger). Product Usage: Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or…
Recalling firm
St. Jude Medical
Quantity
70,638 (combined total of the Eon and the Eon Mini systems)
Distribution
Worldwide Distribution - USA Nationwide Including Puerto Rico and countries of: Argentina, Australia, Belgium, Brazil, Canada, China, Chile, Colombia, Ireland, Israel, India, Malaysia, Mexico, Saudi Arabia, Singapore, South Africa, South Korea, Switzerland and United Kingdom.
Recall date
July 24, 2013
Classification
Class II
Status
Terminated
Agency
FDA (device)