RUSCH Easy Cath Coude Kit recall
RUSCH Easy Cath Coude Kit, 10 FR, Rx Only, Sterile, Distributed by: Teleflex Medical, Research Triangle Park, NC 27708. A tubular device that is inserted through the urethera and is used to pass fluids to or from urinary tracts.
Why was RUSCH Easy Cath Coude Kit recalled?
Product was packaged with a straight catheter rather than the correct coude (curved) catheter.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Product Code: ECK103, Lot number: 02A1402484
- Product
- RUSCH Easy Cath Coude Kit, 10 FR, Rx Only, Sterile, Distributed by: Teleflex Medical, Research Triangle Park, NC 27708. A tubular device that is inserted through the urethera and is used to pass fluids to or from urinary tracts.
- Recalling firm
- Teleflex Medical
- Quantity
- 1000 eaches
- Distribution
- US Distribution including the states of CA, FL, OH and OK.
- Recall date
- June 18, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)