Vitros 5 recall: Software Issue
VITROS 5,1 FS Chemistry System, Software Version 2.8 & Below Automates pre-analytical and post-analytical sample and data management for in vitro…
Why was Vitros 5 recalled?
Software anomaly regarding urine samples that require acidified pretreatment. The software has allowed urine samples for assays that require acidified pretreatment to be metered from the same sample container as assays that required no pretreatment when the sample was programmed.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Product Code: 6801375, 6801890 Unique Device Identifier Numbers: 10758750001132, 10758750001644
- Product
- VITROS 5,1 FS Chemistry System, Software Version 2.8 & Below Automates pre-analytical and post-analytical sample and data management for in vitro…
- Recalling firm
- Ortho-Clinical Diagnostics
- Quantity
- 2,387 units Total (987 domestically & 1400 internationally)
- Distribution
- Worldwide Distribution - US Nationwide, Brazil, Chile, Mexico, Australia, Hong Kong, India, Taiwan, Belgium, Denmark, France, Ireland, Italy, Portugal, Romania, Spain, Sweden and the UK.
- Recall date
- June 17, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)