FDAClass IJune 28, 2023

ICD Ddpa2d4g Cobalt XT DR MRI Df4 Gold recall

ICD DDPA2D4G COBALT XT DR MRI DF4 GOLD, Model Number DDPA2D4G; Implantable Cardioverter Defibrillators

Why was ICD Ddpa2d4g Cobalt XT DR MRI Df4 Gold recalled?

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

GTIN 00763000615628, Lot Serial Numbers: RSM616778S, RSM610606S, RSM611337S, RSM610608S, RSM611339S; GTIN 00763000645502, Lot Serial Numbers: RSM616777S

Product
ICD DDPA2D4G COBALT XT DR MRI DF4 GOLD, Model Number DDPA2D4G; Implantable Cardioverter Defibrillators
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Quantity
6 units
Distribution
US Nationwide - Worldwide Distribution
Recall date
June 28, 2023
Classification
Class I
Status
Ongoing
Agency
FDA (device)