FDAClass IIMay 14, 2025

Washer Disinfector Aquadis 56 recall

Washer Disinfector Aquadis 56; Model Number: 56A

Why was Washer Disinfector Aquadis 56 recalled?

The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model: 56A; UDI-DI: 07340153710344; Serial Numbers: WAA104857, WAA107934, WAA107937, WAA108359, WAA108540, WAA108963, WAA109338, WAA110115, WAA108201, WAA109175, WAA109884, WAA109989, WAA109998, WAA110804, WAA600035, WAA109641, WAA109651, WAA111118, WAA106820, WAA106835, WAA106841, WAA107693, WAA107428, WAA107429, WAA600055, WAA600060, WAA600061, WAA106607, WAA106668, WAA106753, WAA107358, WAA109804, WAA106793, WAA108419, WAA110679, WAA109022, WAA109030, WAA109564, WAA109566, WAA109595, WAA109598, WAA105283, WAA107346, WAA600188, WAA107314, WAA107330, WAA110062, WAA110075, WAA110084, WAA110104, WAA110163, WAA110302, WAA110201, WAA600558, WAA110953, WAA110963, WAA110138, WAA109975, WAA107281, WAA107303, WAA108972, WAA110228, WAA110162, WAA110727, WAA110738, WAA110742, WAA110803, WAA108669, WAA108670, WAA108673, WAA108722, WAA110762, WAA110631, WAA110636, WAA600167, WAA600168, WAA110482, WAA111159, WAA111170, WAA111183, WAA105284, WAA108615, WAA109514, WAA110174, WAA110178, WAA110898, WAA110907, WAA600077, WAA600177, WAA600338, WAA600339, WAA600340, WAA600341, WAA600342, WAA600343, WAA600344, WAA600394, WAA600705, WAA600706, WAA600711, WAA600712, WAA600713.

Product
Washer Disinfector Aquadis 56; Model Number: 56A
Recalling firm
Getinge Disinfection Ab
Quantity
102 units (8 US, 94 OUS)
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Brazil, Canada, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Italy, Japan, Kenya, Korea, Kuwait, Latvia, Lithuania, Malaysia, Mauritania, Mexico, Netherlands, New Caledonia, Nigeria, Norway, Oman, Pakistan, Philippines, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Vietnam, Zambia.
Recall date
May 14, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls