UniCel DxC 600i SYNCHRON Access Clinical System recall
UniCel DxC 600i SYNCHRON Access Clinical System, Part No. A25637, A25638 Product Usage: The Access 2 and UniCel DxI Access Immunoassay Systems are in…
Why was UniCel DxC 600i SYNCHRON Access Clinical System recalled?
Beckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is running. The position errors cause the instrument to stop running and cancel tests that are in process.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial No. (Part No. A25637) 800999,801019 Serial No. (Part No. A25638)800995,800996,800997,800998,801000,801001,801002,801003,801004, 801005,801006,801007,801008,801009,801010,801011,801012,801013, 801014,801015,801016,801017,801018,801020,801021,801022,801023, 801024,801025,801026,801027
- Product
- UniCel DxC 600i SYNCHRON Access Clinical System, Part No. A25637, A25638 Product Usage: The Access 2 and UniCel DxI Access Immunoassay Systems are in…
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 851 units total for all products (291 units in US)
- Distribution
- Affected geographies: Algeria, Austria, Australia, Bangladesh, Bulgaria, Brazil, Canada, China, Costa Rica, Cyprus, Colombia, Czech Republic, El Salvador, France, Germany, Ghana, Greece, Guatemala, India, Indonesia, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Lithuania, Malawi, Mexico, Morocco, Myanmar, Nepal, Oman, Panama, Philippines, Pakistan, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Switzerland, Tunisia, Turkey, United Kingdom, United States, Vietnam, Zimbabwe
- Recall date
- April 12, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)