FDAClass IIApril 22, 2020

Positron Emission Tomograpy and Computed Tomography recall: Software Issue

Positron Emission Tomograpy and Computed Tomography System, Model: uMI 550, with software version R001.3.0.0.750505, Rx Only, CE 0197, UDI…

Why was Positron Emission Tomograpy and Computed Tomography recalled?

Potential sporadic software bugs in R001.3.0.0.750505 software version may cause an issue with the PET acquisition raw data causing reconstruction problems during PET/CT scanning and could possibly cause rescans of patients with additional dose.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers: 200017 and 200023

Product
Positron Emission Tomograpy and Computed Tomography System, Model: uMI 550, with software version R001.3.0.0.750505, Rx Only, CE 0197, UDI…
Recalling firm
Shanghai United Imaging Healthcare Co., Ltd.
Quantity
2 units
Distribution
US Nationwide distribution in the state TX.
Recall date
April 22, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls