Airway Exam Kit recall: Foreign Material
Airway Exam Kit, DYKE1796
Why was Airway Exam Kit recalled?
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI-DI 10193489874716 (ea) 40193489874717 (case) Lots 24CBT023 24FBB701 24LBL343 25GBP141
- Product
- Airway Exam Kit, DYKE1796
- Recalling firm
- Medline Industries, LP
- Quantity
- 14,379 kits total
- Distribution
- US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.
- Recall date
- April 8, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)