FDAClass IIApril 22, 2020

Merlin PCS 3650 Software Upgrade Kit recall: Software Issue

Merlin PCS 3650 Software Upgrade Kit, REF 3330 St. Jude Medical The MerlinTM Patient Care System (MerlinTM PCS) Model 3650 is a portable, dedicated…

Why was Merlin PCS 3650 Software Upgrade Kit recalled?

Internal testing of software identified software anomaly scenarios where if non-applicable shock energy value is selected, the device will continue to use the previously programmed energy level and the user will not be alerted that the selected value was not retained.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software Version: 24.6.1 Model: 3650, GTIN: 05414734500661 Model: 3330, GTIN: 05414734509725 For affected serial number of programmers see Attachment B.1 Programmers For models of ICDs interrogated by affected programmers see Attachment B.2 Devices.

Product
Merlin PCS 3650 Software Upgrade Kit, REF 3330 St. Jude Medical The MerlinTM Patient Care System (MerlinTM PCS) Model 3650 is a portable, dedicated…
Recalling firm
Abbott
Quantity
Software Version 24.6.1 installed into 4733 programmers
Distribution
U.S. AK, Al, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. OUS: None - the firm stated that there are no OUS consignees associated to ICDs that were interrogated from January 22, 2020 to February 6, 2020.
Recall date
April 22, 2020
Classification
Class II
Status
Terminated