FDAClass IIJune 9, 2021

T25 Cannulated Screwdrivers recall

T25 Cannulated Screwdrivers (CSD 025) packaged in Guide Wire Kits (GWK 300) - Product Usage: Intended to be used to insert screws with T25 driving features, which includes screws with diameters of 5.5mm, 6.5mm, and 7.3mm Product Code: GWK 300

Why was T25 Cannulated Screwdrivers recalled?

Cannulated Screwdrivers may break at the driver tip along the shaft, near the handle extending surgery time and additional x-ray to identify the fragments from the breakages

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Numbers: 2003320898, 1810220041, 1802220008, 1802220005, 1711220035 UDI: 00840118110853

Product
T25 Cannulated Screwdrivers (CSD 025) packaged in Guide Wire Kits (GWK 300) - Product Usage: Intended to be used to insert screws with T25 driving features, which includes screws with diameters of 5.5mm, 6.5mm, and 7.3mm Product Code: GWK 300
Recalling firm
Flower Orthopedics Corporation
Quantity
47 units
Distribution
US Nationwide distribution in the states of AZ, CO, FL, GA, IL, IN, PA, NC, VA.