FDAClass IIJune 12, 2019

CyberKnife VSI recall

CyberKnife VSI, Part Numbers: a) 032500-010: CyberKnife VSI 9.6.x Robotic Radiosurgery System for Japan b) 033000-010: CyberKnife VSI Robotic…

Why was CyberKnife VSI recalled?

There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

a) 032500-010: CyberKnife VSI 9.6.x Robotic Radiosurgery System for Japan Serial Numbers: C0281, C0307 b) 033000-010: CyberKnife VSI Robotic Radiosurgery System Serial Numbers: C0042, C0064, C0071, C0074, C0079, C0091, C0098, C0102, C0137, C0161, C0180, C0182, C0187, C0200, C0203, C0210, C0216, C0230, C0238, C0246, C0247, C0249, C0252, C0255, C0257, C0264, C0269, C0270, C0273, C0274, C0276, C0277, C0284, C0293, C0294, C0304, C0305, C0322, C0323, C0325, C0326

Product
CyberKnife VSI, Part Numbers: a) 032500-010: CyberKnife VSI 9.6.x Robotic Radiosurgery System for Japan b) 033000-010: CyberKnife VSI Robotic…
Recalling firm
Accuray Incorporated
Quantity
43 units
Distribution
US & Italy, United Kingdom, Switzerland, Belgium, Canada, Germany, France, Japan, Poland, Netherlands, Luxembourg, Australia, Qatar, Latvia, Portugal, India, Hungary, Turkey, United Arab Emirates, Thailand
Recall date
June 12, 2019
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls