FDAClass IIApril 5, 2017

Baxter SIGMA Spectrum Infusion Pump with Master Drug recall: Software Issue

Baxter SIGMA Spectrum Infusion Pump with Master Drug Library, Model #/Product Code 35700BAX2

Why was Baxter SIGMA Spectrum Infusion Pump with Master Drug recalled?

SIGMA Spectrum Infusion Pumps with Master Drug Library, Software V8.00.02 contains an anomaly that may cause a Sharp Watchdog Timeout error message. Specifically, the software anomaly causes the error message only under conditions in which a message is being written to the Spectrum Infusion Pumps Event History Log (EHL), and the length of that message exactly matches the size of the available memory at the end of a memory sector. For this reason, the issue is significantly more likely to manifest when poor network maintenance results in high volumes of messages being written to the EHL.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

14,362 Serial Numbers (please contact CDRH for list of affected Serial Numbers).

Product
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library, Model #/Product Code 35700BAX2
Recalling firm
Baxter Healthcare Corporation
Quantity
Domestic: 14,175 units; Canada: 187 units
Distribution
Natiowide, Puerto Rico and Canada
Recall date
April 5, 2017
Classification
Class II
Status
Terminated