Semi-Automated External Defibrillators recall: Labeling Error
Semi-Automated External Defibrillators; Models: DDU-100 and DDU-100E. Item numbers: DCF-E110SG-DE/1 (German language), DCF-E110SG-FR/1 (French…
Why was Semi-Automated External Defibrillators recalled?
It was noticed there was no Italian translation included for the AED. Article 16 of the Swiss Medical Devices Ordinance (MedDO) states that product information (labelling and instructions for use) should be provided in all three official languages of Switzerland (German, French, and Italian).
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI-DIs: 00815098020034 (DDU-100, export version), 10815098020031 (DDU-100, export version (configuration level)). Lot numbers: 114008193, 114007350, 114007323, 114008190, 116012370, 116011831, 114006014, 114006766, 114006943, 114007067, 114007068, 114007071, 114007078, 114007091, 114007114, 114007123, 114007128, 114007137, 114007145, 114007478, 114007486, 114007517, 114007623, 114007696, 114007705, 114007770, 114007906, 114007907, 114007911, 114007914, 114007925, 114007933, 114007934, 114007964, 116011703, 116011708, 116011883, 116012476, 116012485, 116012488, 116012501, 104016266, 104016297,
- Product
- Semi-Automated External Defibrillators; Models: DDU-100 and DDU-100E. Item numbers: DCF-E110SG-DE/1 (German language), DCF-E110SG-FR/1 (French…
- Recalling firm
- Defibtech, LLC
- Quantity
- 43 units (OUS only)
- Distribution
- International Only: Switzerland.
- Recall date
- May 7, 2025
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)