FDAClass IIApril 15, 2020

"hemoglobin A1c Calibrators *** recall

"HEMOGLOBIN A1c CALIBRATORS *** REF 4P52-01 *** 307261/R03 *** S4P5X0 *** FOR USE WITH ARCHITECT" - Product Usage: use in the calibration of the…

Why was "hemoglobin A1c Calibrators *** recalled?

Calibrator level 2 bottle may have been manufactured incorrectly for a portion of the lot. The manufacturer observed no impact to assay performance between the HbA1c concentration range of 5.33% to 6.87%. However, outside of this range, a bias of greater than 3% may be seen with patient results. The device failure that is associated with this recall can result in potentially negatively and positively biased patient test results for HbA1c. The firm is investigating the root cause.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

List Number: 4P52-01 Lot Number: 54582UQ02

Product
"HEMOGLOBIN A1c CALIBRATORS *** REF 4P52-01 *** 307261/R03 *** S4P5X0 *** FOR USE WITH ARCHITECT" - Product Usage: use in the calibration of the…
Recalling firm
Abbott Laboratories
Quantity
198 Kits
Distribution
Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, FKL GA, IA, ID, IL, IN, KS, LA, MN, MO, NC, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT, VA, and WI; and countries of Austria, Germany, Italy, South Africa, and Spain.