Iceseed 1.5 CX S Needle US recall
IceSeed 1.5 CX S NEEDLE US, Cryoablation Needle, REF H7493967233100. The needles are intended to convert high-pressure gas to either a very cold…
Why was Iceseed 1.5 CX S Needle US recalled?
Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
GTIN 00191506032678, Batch Numbers: 33744602, 33744603, 33744604, 33744605, 33931901, 33937786, 33937787, 33942987, 33942988, 33942989, 33950417, 33950418, 33996423, 33996424, 33996428, 34038291, 34038292, 34038307, 34117281, 34128141, 34128142, 34151877, 34151878, 34204928, 34204929, 34256930, 34256931, 34256932, 34295233.
- Product
- IceSeed 1.5 CX S NEEDLE US, Cryoablation Needle, REF H7493967233100. The needles are intended to convert high-pressure gas to either a very cold…
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 435 units
- Distribution
- Nationwide
- Recall date
- May 7, 2025
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)