FDAClass IIApril 15, 2020

PRIMUS HI #04504200 with Digital LINAC Systems Control recall

PRIMUS HI #04504200 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13

Why was PRIMUS HI #04504200 with Digital LINAC Systems Control recalled?

Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers 3025, 3049, 3241, 3267, 3305, 3422, 3479, 3515, 3519, 3547, 3565, 3602, 3731, 3751, 3832, 3667, 3924, 3949, 4036, 70-4185, 5153, 3361, 3979, 3657, 3679

Product
PRIMUS HI #04504200 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
230,001
Distribution
Worldwide distribution. US consignees in the following states: Alabama, California, Delaware, Florida, Georgia, Iowa, Illinois, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio, Oklahoma, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Wisconsin, West Virginia, Wyoming
Recall date
April 15, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls