***REF C7120***APEX Arthroscopy Tubing Set. Intended recall
***REF C7120***APEX Arthroscopy Tubing Set. Intended for use for joint irrigation and dissention during arthroscopic procedures.
Why was ***REF C7120***APEX Arthroscopy Tubing Set. Intended recalled?
Conmed Linvatec received complaints that their products C7120 APEX Arthroscopy Tubing Set, One Connection and C7122 APEX Arthroscopy Tubing Set, Two Connection, contained a scratch/grooved area on the inside diameter of the pressure fitting that may cause the APEX Irrigation (pump) Console to sense a pressure drop and increase fluid flow to the surgical site.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
C7120 Lot's #: 1301184, 1301214, 1301254, 1301284, 1301294, 1302014, 1302044, 1302054, 1302064, 1302074, 13020124, 1302134, 1302144, 1302194, 1302204, 1302214, 1302224, 1302254, 1302264, 1303014, 1303054, 1303064, 1303074, 1303114, 1303124, 1303204, 1303214, 1303264, 1303274, 1303284, 1304014, 1304034, 1304054, 1304154, 1304294, 1304304, and 1305014.
- Product
- ***REF C7120***APEX Arthroscopy Tubing Set. Intended for use for joint irrigation and dissention during arthroscopic procedures.
- Recalling firm
- Linvatec Corp. dba ConMed Linvatec
- Quantity
- 4,691 Units total of each
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AR, CA, FL, GA, IL, IN, KS, MO, NC, NJ, NM, NY, OH, OK, PA, TN, TX, UT, VA, and WI, and the countries of Austria; Belgium, Canada, China, Czech Republic, Finland, France, Great Britain, Italy, and Japan.
- Recall date
- July 17, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)