FDAClass IIJune 11, 2014

115 V Blanketrol II recall

115 V Blanketrol II, Model 222S Hyper-Hypothermia System . Used to lower or to raise a patients temperature and /or maintain a desired patient…

Why was 115 V Blanketrol II recalled?

The incorrect voltage component was placed into the device during manufacturing. This may make the device unable to maintain patient temperature during cooling mode. The device will continue to indicate that the unit is cooling because the compressor will continue to run, however water in the reservoir will begin to return to ambient temperature. The actual water temperature will be displayed but

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model Number 222S; Serial #s: 133-2-00697, 133-2-00698; 133-2-00699; 133-2-00700; 133-2-00701; 133-2-00702; and 133-2-00703:.Part # 86165

Product
115 V Blanketrol II, Model 222S Hyper-Hypothermia System . Used to lower or to raise a patients temperature and /or maintain a desired patient…
Recalling firm
Cincinnati Sub-Zero Products Inc
Quantity
7 units
Distribution
The product was distributed to one customer in MI, and internationally (Taiwan)
Recall date
June 11, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls