FDAClass IIMay 16, 2018

Immulite ¿ 2000/immulite ¿ 2000 Xpi Anti Hbc recall

IMMULITE ¿ 2000/IMMULITE ¿ 2000 XPi Anti HBc

Why was Immulite ¿ 2000/immulite ¿ 2000 Xpi Anti Hbc recalled?

Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Anti HBc, Anti HBc (IMMULITE 2000/IMMULITE 2000 XPi L2KHC2 (D)) Lot codes: Anti HBcM # 579 580 581 D579 D580 D581 ***Added 9/11/20*** D602

Product
IMMULITE ¿ 2000/IMMULITE ¿ 2000 XPi Anti HBc
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
N/A
Distribution
Nationwide and Canada, Mexico, and Thailand
Recall date
May 16, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls