Dimension Vista¿ Thyroid Stimulating Hormone recall
Dimension Vista¿ Thyroid Stimulating Hormone (TSH) Flex¿ reagent cartridge
Why was Dimension Vista¿ Thyroid Stimulating Hormone recalled?
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
TSH, radioimmunoassay, thyroid stimulating hormone, RF612, SMN# 10464524 Lot codes: 16277BD 16322BB 17005BB 17023AC 17069AA 17093BB 17122AA 17150AA 17193BA 17215AC 17268AA
- Product
- Dimension Vista¿ Thyroid Stimulating Hormone (TSH) Flex¿ reagent cartridge
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- N/A
- Distribution
- Nationwide and Canada, Mexico, and Thailand
- Recall date
- May 16, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)