FDAClass IIMay 16, 2018

Dimension¿ Free Thyroxine recall

Dimension¿ Free Thyroxine (FT4L) Flex¿ reagent cartridge

Why was Dimension¿ Free Thyroxine recalled?

The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

FT4L, radioimmunoassay, free thyroxime, RF610, SMN# 10464523 Lot codes: EA7276 EC7304 EC7333 EA7346 EB8010 EA8037 EA8065 EB8086 EB8108 EA8130 EA8156 EA8179 EA8205 EA8233 FD8254

Product
Dimension¿ Free Thyroxine (FT4L) Flex¿ reagent cartridge
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
N/A
Distribution
Nationwide and Canada, Mexico, and Thailand
Recall date
May 16, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)