Dimension¿ Thyroid Stimulating Hormone recall
Dimension¿ Thyroid Stimulating Hormone (TSHL) Flex¿ reagent cartridge
Why was Dimension¿ Thyroid Stimulating Hormone recalled?
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
TSHL, radioimmunoassay, thyroid-stimulating hormone, RF612, SMN# 10464524 Lot codes: EB7276 EC7298 EA7318 EB7346 EC7346 EA8012 EA8033 EA8051 EC8080 EB8107 EB8121 EA8143 EB8143 EA8177 EB8177 EB8199 EA8223 EA8241 EA8261
- Product
- Dimension¿ Thyroid Stimulating Hormone (TSHL) Flex¿ reagent cartridge
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- N/A
- Distribution
- Nationwide and Canada, Mexico, and Thailand
- Recall date
- May 16, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)