FDAClass IIMay 25, 2016

Anspach recall

ANSPACH, Rx only, Otologic Curved Micro (OCCM) Diamond Ball Burr, 1.5 MM, Part # OCM7-15SD Product Usage: The Otologic Curved Micro (OCCM) Attachment…

Why was Anspach recalled?

Certain lots of the Otologic Curved Micro (OCM) Diamond Ball Burrs may be missing the diamond coating and the ball may be the incorrect size.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model/Catalog #:OCM7-15SD, 1.5 MM Diamond Ball, Lot Numbers: J473111704 and J473111705

Product
ANSPACH, Rx only, Otologic Curved Micro (OCCM) Diamond Ball Burr, 1.5 MM, Part # OCM7-15SD Product Usage: The Otologic Curved Micro (OCCM) Attachment…
Recalling firm
The Anspach Effort, Inc.
Quantity
27 devices
Distribution
US Distribution in the states of KY and TX.
Recall date
May 25, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls