FDAClass IIApril 8, 2026

RefleXion X1 Radiotherapy System recall: Software Issue

RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00

Why was RefleXion X1 Radiotherapy System recalled?

Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software Version: 2.1.29-4 and Software Version: 2.1.35-1. UDI: (01)00860003983812(11)230504(21)X11001, (01)00860003983812(11)230518(21)X11002, (01)00860003983812(11)230505(21)X11003, (01)00860003983812(11)230530(21)X11004, (01)00860003983812(11)230811(21)X11006, (01)00860003983812(11)230410(21)X11007, (01)00860003983812(11)230803(21)X11008, (01)00860003983812(11)240528(21)X11009, (01)00860003983812(11)240820(21)X11010, (01)00860003983812(11)241107(21)X11011, (01)00860003983812(11)241206(21)X11012, (01)00860003983812(11)250731(21)X11013.

Product
RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00
Recalling firm
Reflexion Medical, Inc.
Quantity
12
Distribution
US: CA, TX, PA, CT, NJ, OR, LA, OH
Recall date
April 8, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls