Alaris Infusion Pump Module recall
Alaris Infusion Pump Module, Model 8100, serviced/repaired by The Biomed Guys, with the impacted bezel repair part, TIPA-8100-4410 / Alaris 8100…
Why was Alaris Infusion Pump Module recalled?
Bezel repair part, not by the original manufacturer, used in the service and repair of Alaris Pump Module Model 8100, may experience cracking and/or separation of the bezel posts. Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial numbers of serviced devices, serviced 09/2020-02-2021: 12833065, 13187021, 13187811, 12838069, 12982509, 12837231, 12835186, 12837972, 12710593, 12776790, 13753893, 13605919, 12832544, 15763526, 14806203, 14806612, 14807070, 13738432, 15763547, 15865626, 14806472, 12796415,
- Product
- Alaris Infusion Pump Module, Model 8100, serviced/repaired by The Biomed Guys, with the impacted bezel repair part, TIPA-8100-4410 / Alaris 8100…
- Recalling firm
- The Biomed Guys
- Quantity
- 22 units
- Distribution
- Us distribution to the states of IN and TX.
- Recall date
- June 9, 2021
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)