InPen smart insulin pen recall
InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA
Why was InPen smart insulin pen recalled?
Insulin pens may have been incorrectly assembled therefore users could experience the following: either the insulin cartridge will not fit into the cartridge holder; or the cartridge holder may be difficult to remove from the insulin pen. If the issue is present, and if the user does not have backup insulin they may experience temporary hyperglycemia.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
REF/UDI-DI(GTIN)/Lot: MMT-105NNBLNA/763000827182/D0020, D0034, D0036, D0038, D0045, D0046, D0050, D0057; MMT-105NNGYNA/763000827199/D0026, D0037, D0039, D0041, D0042, D0043, D0049; MMT-105NNPKNA/763000827205/D0029, D0030, D0040, D0051, D0062; MMT-105ELBLNA/763000827151/D0021, D0033, D0035, D0044, D0059, D0060, D0061; MMT-105ELGYNA/763000827168/D0031, D0032, D0047, D0068; MMT105ELPKNA/763000827175/D0027, D0028, D0048
- Product
- InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA
- Recalling firm
- Medtronic MiniMed, Inc.
- Quantity
- 8,986
- Distribution
- US: OK, IA, NC, SC, MI, MN, CA, AZ, AL, MO, WA, IL, KS, TX, FL, MT, MD, OH, VA, NY, TN, HI, MA, NJ, PA, CO, WI, GA, CT, NE, OR, LA, NM, SD, ND, IN, NV, WV, RI, AR, NH
- Recall date
- April 30, 2025
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)